Edt calibration command and command data summary – Rockwell Automation 1794-IRT8XT FLEX I/O Thermocouple/RTD /Millivolt Input Module User Manual User Manual

Page 56

Advertising
background image

Publication 1794-6.5.12 - September 2011

44

Calibrate Your Module

EDT Calibration Command and Command Data Summary

Config Word 3
EDT Command
bits 8…15

Meaning

Config Word 3
bits 4…7

Config Word 3
bits 0…3

Meaning

Dec. (Hex)

Channel No.

Command data

4 (4)

General calibration

by channel

0…7

1

Channel internal current source

and current sense resistor, with

4-wire external 383

Ω

inputs

0…7

2

Channel internal current source

and current sense resistor with

4-wire external 100

Ω

inputs

0…7

3

Loop channel compensation for

2-wire RTD

0…7

4

CJC calibration

5 (5)

Offset calibration

by channel

0…7

0

gain= 1, input = -320 mV

0…7

1

gain = 2, input = 0.0 mV

0…7

2

gain = 4, input = -50 mV

0…7

3

gain = 8, input = -10.0 mV

0…7

4

gain = 16, input = -9 mV

0…7

5

gain = 32, input = 1.0 mV

6 (6)

Gain calibration by

channel

0

gain = 1, input = 320 mV

1

gain = 2, input = 320 mV

2

gain = 4, input = 110 mV

3

gain = 8, input = 70 mV

4

gain = 16, input = 29 mV

5

gain = 32, input = 19 mV

36 (24)

Calibration all

channels
General calibration

all channels

0

1

Channel internal current source

and current sense resistor, with

4-wire external 383

Ω

inputs

0

2

Channel internal current source

and current sense resistor with

4-wire external 100

Ω

inputs

0

3

Loop channel compensation for

2-wire RTD

0

4

CJC calibration

Offset calibration

all channels

1

0

gain = 1, input = -320 mV