Auto setup process – Rockwell Automation 5370-CBPK Color CVIM BLISTER-PAK Option User Manual

Page 29

Advertising
background image

Chapter 5
Auto Setup

5–9

The auto setup process begins when you pick

Auto Setup

in the

Blister

Setup

menu bar. It involves the following main tasks:

Focusing the camera and obtaining an image of good blisters.

Positioning an image window around all blisters.

Teaching the background color.

Teaching the pocket color (for aluminum blisters).

Teaching the tablet or capsule color (or colors).

Positioning a window around one blister, and automatically creating
“clones” of the window around all other blisters.

Setting the minimum object area.

Performing the “teach–in” function, during which the system calculates
tolerance parameters for position, area, and shape variations.

When you pick the

Auto Setup

menu item, the

Auto Setup

menu appears,

as shown in Figure 5.2.

Figure 5.2
Selecting Auto Setup Menu

The

Focus Camera

option enables you to obtain a setup image. The lower

three menu items in the

Auto Setup

menu identify the three entry points to

the auto setup process, as follows:

Learn Colors

–– From this entry point, the system will step through the

complete auto setup process and create a complete new blister
configuration.

Position Windows

–– This entry point bypasses the

Learn Colors

steps

and goes directly to the point of positioning a teach window. Accessing

Position Windows

would be appropriate when the size and/or position of

any of the blister windows must be changed, but the colors do not need to
be re–learned.

Teach–in

–– This entry point bypasses the

Learn Colors

and

Position

Windows

steps and goes directly to the point of learning the object

positions. Accessing

Teach–in

would be appropriate when the nominal

object positions must be re–learned and/or a new set of samples
evaluated, but neither the colors nor the blister window positions need to
be changed.

Auto Setup Process