Bio-Rad BioFrac™ Fraction Collector User Manual
Page 30

3
| System Configuration and Plumbing
22
| BioFrac Fraction Collector
collector to another manufacturer’s pump, UV monitor, and/or chart recorder, you
need a BioFrac accessory cable (15-pin to bare wires), system cable 7 (mini-DIN to
bare wires), and 2-wire shielded cable (26 gauge or larger, available from most
electronics or hardware stores). In addition, the pump and chart recorder circuit
logic must be Start/closed circuit = ON; Stop/open circuit = OFF. The UV monitor’s
analog output must be 100 mV or 1 V. (Refer to
Appendix B, Rear Panel Connector
91), for details.)
Fig. 10.
Connecting the fraction collector to components other than Bio-Rad’s.
To remotely start or stop an external pump from the BioFrac fraction
collector
1.
Chapter 5, Stand-Alone Operation
collector to Local mode. In Local mode, all fraction collection parameters are
programmed on the BioFrac device’s front panel.
2.
Connect the BioFrac accessory cable (15-pin to bare wires; see
Rear Panel Connector Information
85) to the I/O port on the back of
the fraction collector.