Bio-Rad BioFrac™ Fraction Collector User Manual

Page 30

Advertising
background image

3

| System Configuration and Plumbing

22

| BioFrac Fraction Collector

collector to another manufacturer’s pump, UV monitor, and/or chart recorder, you

need a BioFrac accessory cable (15-pin to bare wires), system cable 7 (mini-DIN to

bare wires), and 2-wire shielded cable (26 gauge or larger, available from most

electronics or hardware stores). In addition, the pump and chart recorder circuit

logic must be Start/closed circuit = ON; Stop/open circuit = OFF. The UV monitor’s

analog output must be 100 mV or 1 V. (Refer to

Appendix B, Rear Panel Connector

Information

(page

85), and

Appendix C, Specifications

(page

91), for details.)

Fig. 10.

Connecting the fraction collector to components other than Bio-Rad’s.

To remotely start or stop an external pump from the BioFrac fraction
collector

1.

As described in

Chapter 5, Stand-Alone Operation

(page

57), set the fraction

collector to Local mode. In Local mode, all fraction collection parameters are
programmed on the BioFrac device’s front panel.

2.

Connect the BioFrac accessory cable (15-pin to bare wires; see

Appendix B,

Rear Panel Connector Information

, on page

85) to the I/O port on the back of

the fraction collector.