5

| Stand-Alone Operation

64

| BioFrac Fraction Collector

number and end tube number are automatically updated according to the
MultiRun mode selected.

6.

Set the fraction size units (min, ml, or drops) and set the fraction size. If
collecting by volume, you must enter a flow rate.

7.

Press Collect (F4) and turn Windows on.

8.

Press Table (F3) and enter the windows and fraction size in the Collection
Windows Table screen. Note that the fraction size entered in the Main screen is
the default size. The fraction size can be changed for each window. When
finished, press Done. On the Main screen, Collect should now read Collect –>
Windows.

9.

(Optional) press Collect (F4) and set the delay time (see

Collect Using a Delay

Function on page

67

).

10. (Optional) press Method (F3) and save your method.

11. Press Run (F1) to start the experiment.

Note:

At any point during the run you can use the Run screen function

keys (see

Table

8 on page

43

).