Bio-Rad BioFrac™ Fraction Collector User Manual

Page 38

Advertising
background image

3

| System Configuration and Plumbing

30

| BioFrac Fraction Collector

Ice Bath

Remove the lid of the ice bath/microplate rack and fill the tub approximately ½ full

with crushed ice, replace the lid and insert 13 x 100 mm culture tubes. Remove the

rack positioning tray from the fraction collector and replace it with the ice bath rack.

Adjust the fraction collector control module appropriately for the height of the tubes

being used. The height adjustment lock pins (see

Figure

11 on page 32

) can be used

to lock the drophead at the desired height. Tube number 1 is located in the front, left

corner of the rack.

Microplates and Titertube Tubes (Racks P1, P2, P3, P4, and TT1)

Remove the rack positioning tray from the fraction collector and replace it with the

ice bath/microplate rack. Position the ice bath/microplate rack such that rack tube

number 1 is in the left, front corner. Mount the plates on the rack using the plate

positioning tabs located on the top of the rack. Adjust the fraction collector control

module appropriately for the height of the plates being used. The height adjustment

lock pins (see

Figure

11 on page 32

) can be used to lock the drophead at the

desired height. Position the plates such that tube A1 is in the left, front corner. In

order to obtain a uniform fraction size when collecting fractions smaller than 0.5 ml,

we strongly suggest that fraction size be specified in drops rather than time or

volume. Alternatively, the microplate Drop Head Kit (catalog # 741-0088) can be

used to obtain finer resolution of fraction sizes since it delivers an approximately

25

l drop.

Preparative (Prep-20) Adaptor

Attach Tygon tubing to each funnel that is long enough to reach the collection

vessels (20 feet of 3/8" OD, 1/4" ID Tygon tubing is provided for this purpose.)

Ensure that no kinks constrict flow in the tubing. Gravity flow from the prep adaptor

requires that the container used for collection be mounted below the Prep-20

adaptor funnels. A drain trough is provided as added security in the event the tubing

becomes plugged or kinked, preventing sample loss. The drain trough tubing should

be inserted into a clean empty collection vessel.

Advertising