3-Pin Combicon Connector

User Guide

| 89

3-Pin Combicon Connector

5

PEN UP/DOWN

Level

Low

Brown

Chart recorder control,
pen up/down. Active in
Local mode.

6, 7

ADS TERMINAL

Level

Low

Purple blue

Event marks for Method
Start and Fraction
Advance by way of a
relay. Pins #6 and #7 are
closed for 200 ms with
each fraction advance.
Active in Local mode.

8

SIGNAL GROUND

Green

Signal ground.

All input and output pins, except the relay terminals, are TTL compatible. All input pins are with
10 k

 pull-up. The relay pins are used to control low-voltage DC contact-closure type devices,

such as 12 V, 10 mA.

Caution: Do not apply 110 VAC or 220 VAC directly to the relay terminal pins.

Table 25.

8-Pin mini DIN connector output signals, continued.

Pin #

Output Signals

Type

Active

Wire Color

Description

Table 26.

3-Pin

Combicon

connector

Pin #

Input Signals

Type

Active

Description

1

ANALOG
GROUND

Common

UV monitor input: analog ground

2

ANALOG IN(–)

Input

UV monitor input 100 mV and 1 V (–)

3

ANALOG IN(+)

Input

UV monitor input 100 mV and 1 V (+)