Pin combicon connector – Bio-Rad BioFrac™ Fraction Collector User Manual
Page 97
3-Pin Combicon Connector
User Guide
| 89
3-Pin Combicon Connector
5
PEN UP/DOWN
Level
Low
Brown
Chart recorder control,
pen up/down. Active in
Local mode.
6, 7
ADS TERMINAL
Level
Low
Purple blue
Event marks for Method
Start and Fraction
Advance by way of a
relay. Pins #6 and #7 are
closed for 200 ms with
each fraction advance.
Active in Local mode.
8
SIGNAL GROUND
Green
Signal ground.
All input and output pins, except the relay terminals, are TTL compatible. All input pins are with
10 k
pull-up. The relay pins are used to control low-voltage DC contact-closure type devices,
such as 12 V, 10 mA.
Caution: Do not apply 110 VAC or 220 VAC directly to the relay terminal pins.
Table 25.
8-Pin mini DIN connector output signals, continued.
Pin #
Output Signals
Type
Active
Wire Color
Description
Table 26.
3-Pin
Combicon
connector
Pin #
Input Signals
Type
Active
Description
1
ANALOG
GROUND
Common
UV monitor input: analog ground
2
ANALOG IN(–)
Input
UV monitor input 100 mV and 1 V (–)
3
ANALOG IN(+)
Input
UV monitor input 100 mV and 1 V (+)