Run Screen

User Guide

| 43

Table 8.

Run screen parameters (Local mode)

Parameter

Function

Status

Displays the current fraction collection status:

Collecting – the diverter valve is in the collect
position and fractions of a user-specified size
are being collected.

Diverting – the diverter valve is in the Waste
position. Run Paused!: The run is currently
paused and the diverter valve is at Waste. A
warning is displayed that the diverter valve is
diverting to Waste.

Collecting Non-Peak – the diverter valve is in
the collect position but is collecting fractions
of non-peak size.

Will Collect at End of Delay – the fraction
collector is waiting for the delay time before
it switches the diverter valve to Collect.

Will Divert at End of Delay – the fraction
collector is waiting for the delay time before
it switches the diverter valve to Waste.

Run Time

Displays the current run time (in hours and
minutes). Stops incrementing during a pause. If
collecting by volume, the run volume is also
displayed.

Cur. Rack # Tube #

Displays the current rack position and tube
number.

Threshold

Displays the current threshold setting as a
percentage of full scale (%AUFS). (Displayed
only if collecting by threshold).

Current UV

Displays the current UV signal value as a
percentage of full scale (%AUFS). (Displayed
only if collecting by threshold).