6 lp/econo mode operation, Chapter 6, Lp/econo mode operation – Bio-Rad BioFrac™ Fraction Collector User Manual

Page 83: Chapter 6, lp/econo, Mode operation, In lp/econo mode, the remote instrument, 6lp/econo mode operation

Advertising
background image

User Guide

| 75

6

LP/Econo Mode Operation

In LP/Econo™ mode all aspects of fraction collection are controlled by an external

controller. In this mode, only Rack, MultiRun, Start Tube, and End Tube parameters

can be set from the fraction collector Main screen. Time and volume windows and

threshold and delay functions are not programmable from the fraction collector in

this mode. However, if the fraction collector is connected to a BioLogic™ LP

chromatography system, Collect All, Threshold, Collection Windows, and Threshold

and Collection Windows (all with delay volume, if desired) can be controlled through

the BioLogic LP system. When the BioFrac™ fraction collector is connected to a

Model EP-1 Econo pump, only Collect All (with delay volume using void (Vo)) is

available. Any controller capable of sending TTL fraction advance and diverter valve

signals can be used in this mode if it has compatible control circuitry logic (see

Appendix B, Rear Panel Connector Information

, on page

85).

To run the fraction collector in LP/Econo mode

1.

Change mode to LP/Econo and ensure the appropriate cable is connected to
the fraction collector. (See

Chapter 3, System Configuration and Plumbing

, on

page

13 for cabling information.) Connect system cable 15 to the I/O port on

the fraction collector and to the BioLogic LP system or Model EP-1 Econo
pump.

2.

Connect system cable 3 to the Rec port on the BioFrac fraction collector and to
the BioLogic LP System or Model EP-1 Econo pump (see

Chapter 3, System

Configuration and Plumbing

, on page

13 for cabling information).

3.

To select a rack, press Rack (F2) and then choose a rack from the menu.
Choose whether you want the diverter valve to divert to waste during fraction

Advertising