Connecting Instruments and Devices to the BioFrac Fraction Collector

User Guide

| 31

To install the Prep-20 adaptor, remove the two green rubber plugs on the left, front

and right, front of the fraction collector. After attaching the Tygon tubing, insert the

Prep-20 adaptor into the preparative rack holder slots. Notice that the drain trough

slopes slightly towards the drain funnel. The collection port tubing extends down the

front of the fraction collector.

Note:

The Prep-20 adaptor is rated for use with flow rates up to 100 ml/min.

(For discussion of the plumbing for high flow rates, see

Chapter 3, System

Configuration and Plumbing

, on page

13.)

Bottles

Place the rack positioning tray, deep side facing up, and place the bottles in the

circles labeled A, B, C, and D. Adjust the fraction collector head appropriately for the

height of the bottles being used.

Adjusting the Drophead Height

The BioFrac fraction collector accommodates collection vessel heights up to 150

mm. Detents, arranged at predefined heights, are positioned along both columns for

added security. The detent positions accommodate all standard tube and

microplate heights. Lock pins fit snugly within the detents to secure the control

module height and prevent the control module from slipping.