Collect all – Bio-Rad BioFrac™ Fraction Collector User Manual
Page 67
Collect All
User Guide
| 59
Collect All
Fig. 23.
Running a Collect All method.
To run a Collect All method
1.
On the Main screen, ensure that Collect is set to Collect –> All. If the screen is
not set to Collect –> All, press Collect (F4) and turn Windows and Threshold off.
Alternatively, press Method (F3) and load the method Default.
2.
To select a rack, press Rack (F2) and choose a rack from the Rack menu.
Choose whether you want the diverter valve to divert to waste during fraction
advances. If you are collecting in microplates or Titertube tubes, choose a
collection pattern: Serpentine, Row, or Column.
Note:
Whenever a new rack is selected, the start tube and end tube
values are updated to reflect the maximum number of tubes available for
the selected rack.
3.
Choose the appropriate MultiRun mode for your experiment (see
for a description of the MultiRun function).
4.
Set the start tube and end tube rack position and tube number.