Collection Windows Table Screen

The Collection Windows Table screen is used to enter the collection windows

parameters (see

Figure

21

,

Table

17

, and

Table

18

). Up to 20 different collection

windows can be defined, each with a different fraction size and threshold. The

threshold parameter column (Thold) shown in

Figure

21

is displayed only when

threshold collection has been turned on in the Advanced Collection screen (see

Advanced Collection Screen on page

51

). Windows start and end parameters are

entered in minutes (when collecting by time or drops) or in milliliters (when collecting

by volume). The menu at the top of the fraction size column can be used to change

the fraction size units. The global fraction size and threshold entered on the Main

screen (see

Main Screen on page

36

) and Advanced Collection screen, respectively,

are the default fraction size and threshold for the first window.

Fig. 21.

Collection Windows Table screen.

Table 17.

Collection Windows Table screen parameters

Parameter

Function

Start

End

The Collection Windows Table start and end
times are in units of time (min) or volume (ml).

The Collection Windows Table start and end
times are in units of time if collecting fractions
by drop.