Bio-Rad BioFrac™ Fraction Collector User Manual

Page 90

Advertising
background image

7

| Maintenance and Troubleshooting

82

| BioFrac Fraction Collector

Fraction collector
causes a high
backpressure

Tubing is kinked.

Wrong tubing is being
used.

Divert valve or tubing is
plugged.

Replace tubing.

Connect the appropriate size tubing for the
flow rate being used (see

Chapter 3,

System Configuration and Plumbing

,

on

page

13

). Rinse the diverter valve with

water to remove salt crystals or other
particulate matter.

Fraction collector will
not go to system mode
when connected to a
BioLogic DuoFlow

chromatography system

BioFrac system is not
in Main screen.
BioFrac system is not
in Local mode. Bus
cable is not
connected.

Place the fraction collector in Local mode
and make sure the Main screen (start-up
screen) is displayed.

Table 21.

Issues, causes, and solutions, continued.

Issue

Possible Cause

Solution