Specifications
The specifications for the BioFrac™ fraction collector follow in
.
Table 27.
BioFrac fraction collector specifications
Fractionation
Time
0.01 to 9999.9 min
Drop
1 to 99999 drops; flow rate 5.0 ml/min (maximum)
Volume
0.02 to 99999 mls.
Collection methods
Time, drop, time windows and volume windows (up to
20 windows), peak detection, time or volume windows
plus peak detection
Peak detection
Threshold, including slope detection algorithm for
collection of double peaks above a set threshold
Drop counting capability
Up to 5 ml/min flow rates
Maximum collection
volume
Virtually unlimited when used with the Prep-20
adaptor
Tube change time
<350 ms, with rack F1
<350 ms, with rack F2
Detector input
±0 to 100 mV and ±0 to 1 V
Event marker
TTL diverter/peak mark; contact closure fraction mark
Input power
100–240 VAC, ±10%, 50/60 Hz, single phase
Input current
1.2 A (max.); fuse rating inside the power supply is 2 A
Power consumption
45 W
Fusing