Example: collect all with delay – Bio-Rad BioFrac™ Fraction Collector User Manual

Page 78

Advertising
background image

5

| Stand-Alone Operation

70

| BioFrac Fraction Collector

Example: Collect All with Delay

Figure

27

shows an example in which the delay is 0.2 min and the fraction size is

1

min. The diverter valve remains at divert during the delay time. The chart recorder

makes the first tick mark (following the start/run mark) at 1.0 min, and each

subsequent tick mark at 1 min intervals. The drophead advances at 1.2 min and

continues advancing at 1 min intervals. Each fraction size is still 1 min.

Fig. 27.

Example showing Collect All with a delay of 0.2 minutes.

Figure

28

shows an example of a Collect All experiment in Drop mode with a 2-drop

delay and a 10-drop fraction size. In this case the chart recorder makes its first tick

mark (following the start/run mark) at 8 drops:

(fraction size

delay = 10 drops

2 drops = 8 drops)