Rack screen – Bio-Rad BioFrac™ Fraction Collector User Manual
Page 54
4
| The User Interface
46
| BioFrac Fraction Collector
Rack Screen
The Rack screen is used to select the rack type and collection pattern to be used in
,
). The default collection pattern is
serpentine; however, it may be changed to collection by row or by column when
collecting in microplates or Titertube tubes. The Divert Between Tubes feature can
be used to reduce the amount of liquid spilled during tube advances. When this
option is turned on (default), the diverter valve switches to waste during tube
advances. However, if this option is turned off (recommended when collecting in
drop mode at low flow rates), the diverter valve switches to Waste only when the
drophead is moving between nonadjacent tubes.
Fig. 18.
Rack
screen.
Table 11.
Rack
screen
parameters
Parameter
Function
Current Rack
Menu used for rack selection.
Divert Between Tubes
Allows the user to turn off the Divert Between
Tubes function. The fraction collector still
diverts when the drophead is moving between
nonadjacent tubes.
Collection Pattern
Displays the current fraction collection pattern.
This pattern can be changed to row or column
for microplates and Titertube tubes.