Collect by windows and threshold – Bio-Rad BioFrac™ Fraction Collector User Manual
Page 73
Collect by Windows and Threshold
User Guide
| 65
Collect by Windows and Threshold
Fig. 26.
Running a Collect by Windows and Threshold method.
To run a Collect by Windows and Threshold method
1.
To select a rack, press Rack (F2) and then choose a rack from the Rack menu.
Choose whether you want the diverter valve to divert to waste during fraction
advances. If you are collecting in microplates or Titertube tubes, choose a
collection pattern: Serpentine, Row, or Column.
Note:
Whenever a new rack is selected the start tube and end tube values
are updated to reflect the maximum number of tubes available for the
selected rack.
2.
Choose the appropriate MultiRun mode for your experiment (see
for a description of the MultiRun function).
3.
Set the start tube and end tube rack position and tube number. The fraction
collector stops when the last tube is reached unless a stop command is
received before reaching the last tube. At the end of a run, the start tube