Bio-Rad BioFrac™ Fraction Collector User Manual

Page 95

Advertising
background image

15-Pin D Connector

User Guide

| 87

Table 23.

15-Pin D connector output signals

Pin #

Output Signals

Type

Active

Wire Color

Description

4

READY

Level

Low

Green

Indicates that the fraction
collector diverter valve is
connected and functioning
properly.

9

10

START REMOTE

DEVICE

Relay

Closed

Green/Black

Orange/Black

Relay used for starting a
remote device such as a
pump. Relay is closed
when Run is pressed and
opened when Stop is
pressed. Active in Local
mode.

11

EVENT MARK

Pulse

Low

Blue/Black

Sends a negative pulse
(200 ms) for each fraction
advance. If the delay
function is being used, the
drophead advance will be
separated from the event
mark by the delay time,
volume, or drops.

15

SIGNAL

GROUND

Blue/White

Signal ground.

All input and output pins, except the relay terminals, are TTL compatible. All input pins are with
10 k

 pull-up. The relay pins are used to control low-voltage DC contact-closure type devices,

such as 12 V, 10 mA.

Caution: Do not apply 110 VAC or 220 VAC directly to the relay terminal pins.