Bubble Filter Time

The bubble filter time function is used in conjunction with the Peak Collection by

Threshold. The bubble filter time function distinguishes true chromatographic peaks

from unwanted signals such as electrical spikes or the passage of an air bubble.

Such false signals are characterized by extremely fast rise and fall times.

The bubble filter time function is entered at the end of the threshold programming

sequence. The default value is off (bubble filter = 0 seconds).

A signal that exceeds the set threshold value with a rise time of >6.0% full scale

(AUFS) in 0.1 seconds is perceived by the fraction collector as possibly signifying a

false peak. In such cases, the fraction collector looks for a bubble filter time set by

the user. If the signal falls to 2% of the pre-rise FS level within the set bubble filter

time, the signal is rejected and the false peak is not collected. Signals that exceed

the threshold value and are of a longer duration than the bubble filter time are

collected as true peaks.

Very sharp, true chromatographic peaks (such as those from an HPLC application)

may rise at a sufficient rate to trigger this function. As a consequence, if an

unsuitable bubble filter time has been set, then the early part of such peaks may not

be collected.

The bubble filter time can be set from 0 to 10 sec. For most chromatographic

applications, a filter time of 0 or 1 sec will suffice. The actual setting depends upon

both flow rate and peak duration and should be optimized for each type of

separation. Generally speaking, a very short bubble filter time should be used with

sharp peaks of short duration.