Response Mode Control

The Response Mode Control allows you to set the SpO2 measurement time to either

Normal or Fast.

Perfusion Index

Perfusion Index (PI) is an SpO2 feature available only with Masimo-equipped monitors.

PI is a relative reading of pulse strength at the monitoring site. PI display ranges from 0.1

percent (very weak pulse strength) to 20.0 percent (very strong pulse strength). PI is a

relative number and varies between monitoring sites and from patient to patient, as

physiological conditions vary.

During sensor placement, the PI can be used to evaluate the appropriateness of an

application site, looking for the site with the highest PI number. Placing the sensor at the

site with the strongest pulse amplitude (highest PI number) improves performance

during motion. Monitor the trend of the PI for changes in physiological conditions.

SatSeconds™ alarm management

The SatSeconds feature is an SpO2 alarm management system available only with

monitors that are equipped with Nellcor OxiMax Technology.

The SatSeconds feature is the product of the time and magnitude that a patient falls

outside of the SpO2 alarm limits. For example, three points below the alarm limit for 10

seconds equals 30 SatSeconds. An alarm is triggered only when a desaturation event

reaches the SatSeconds limit. The SatSeconds feature is clinician controlled and can be

set to 0, 10, 25, 50, or 100 SatSeconds. If a desaturation event resolves on its own

within the preset time, the clock will automatically reset and the monitor will not alarm.

Note

The SatSeconds feature has a built-in safety protocol that sounds an alarm
whenever three SpO2 violations of any amount or duration occur within a
1-minute period.

144 Patient monitoring

Welch Allyn Connex

®

Vital Signs Monitor 6000 Series™

Advertising