Protection classifications, all monitor configurations

Graphical display resolution

Display area

8 in. (H) x 4 in. (V) (19.5 [H] cm x 11.3 [V] cm)

Pixels

1024 (H) x 600 (V)

Pixel arrangement

RGB (red, green, blue)

Color depth

16 bits per pixel

Speaker volume

Output sound pressure

67 dB at 1.0 meter

Alarm and pulse tones

per IEC 60601-1-8

Pulse frequency (f

0

)

150 – 1000 Hz

Number of harmonic components in the

range 300 Hz to 4000 Hz

minimum of 4

Effective pulse duration (t

d

)

high priority: 75 –200 ms
medium and low priority: 125 – 250 ms

Rise time (t

r

)

10 – 20% of t

d

Fall time

a

(t

f

)

t

f

< t

s

– t

r

Note

The relative sound pressure level of the harmonic components should be within 15 dB above or below
the amplitude at the pulse frequency.

a

Prevents overlap of pulses.

Battery specifications

6 cell

9 cell

Rating

11.1 V 3.80Ah (42Wh)

10.8 V 6.75Ah (73Wh)

Composition

Lithium-ion

Lithium-ion

Charge time to 80 percent capacity

2hr 7m

2hr 25m

Charge time to 100 percent capacity

3hr

4hr

Patient exams per charge

1

26

47

Age to 70 percent capacity

2

300

300

Operating time per charge for continuous
monitoring

3, 4

Do not use

2hr (only if clinician remains in the

room)

196 Specifications

Welch Allyn Connex

®

Vital Signs Monitor 6000 Series™