Maintenance and service

Perform periodic checks

Welch Allyn recommends that each facility conduct periodic checks of each monitor.

1. Check the following at least daily:

Audible indicators (a piezo beeper and a speaker tone) at power-up

Fan, especially at startup

Touchscreen alignment

Date

Time

2. Visually inspect the following at least weekly:

The monitor for any damage or contamination

The AC cord for exposed copper

All cables, cords, and connector ends for damage or contamination

All mechanical parts, including covers, for integrity

All safety-related labeling for legibility and adhesion to the monitor

All accessories (cuffs, tubing, probes, sensors) for wear or damage

Documentation for current revision of the monitor

3. Do the following at least monthly:

Test the speaker by selecting each alarm volume level and listening for the

different tones

Visually inspect the mobile stand wheels for wear and faulty operation

Visually inspect the mounting screws on the wall or cart for looseness and wear

Update settings, replace items, or call for service as necessary based on results of visual

inspection. Do not use the monitor if you see any signs of damage. Qualified service

personnel must check any monitor that is damaged for proper operation before putting

the monitor back into use.

Replace the printer paper

The printer is located on the top of the monitor. Follow these steps to replace the roll of

printer paper:

1. Grasp the two tabs and pull up to open the printer door.