Protection classifications, all monitor configurations
Characteristic
Specification
Electrical rating
100 – 240 V AC, 50 – 60 Hz, 0.8 –1.5 A
Duty cycle
Continuous operation
Type of protection against electric shock
Class I equipment (protectively earthed) with double insulation
Degree of protection against electric
shock, for parts applied to patients
Type BF defibrillator proof
IEC EN 60601-1, 2nd Edition
Recovery time following defibrillator
discharge
Less than or equal to 10 seconds
Flammable anesthetics
WARNING Not suitable for use with flammable
anesthetics.
Degree of protection provided by the
enclosure with respect to harmful ingress
of liquids
IPX0
Non-protected according to EN/IEC 60529; Pulse oximeter equipment
complies with ISO 9919 Cl. 44.6 Ingress of liquids tests and EN/IEC 60601-1,
60601-2-30, 60601-2-49 Cl. 44.3 Spillage tests
Height
Standard chassis: 10 in. (25.4 cm)
Extended chassis: 10 in. (25.4 cm)
Width
Standard chassis: 11 in. (27.9 cm)
Extended chassis: 11.38 in. (28.9 cm)
Depth
Standard chassis: 6 in. (15.3 cm)
Extended chassis: 7.5 in. (19.1 cm)
Weight (including battery)
Standard chassis: 9.5 lb (4.3 kg)
Extended chassis: 10.4 lb (4.7 kg)