Configure sphb alarms, Set sphb averaging mode, Measure sphb – Welch Allyn 6000 Series Connex Vital Signs Monitor - User Manual User Manual
Page 156

1. Touch the Settings tab.
2. Touch the Setup tab.
3. Touch the SpHb vertical tab.
4. Select the desired Trend period.
5. Touch the Home tab.
The new settings take effect immediately.
Configure SpHb alarms
Follow these steps to set alarm limits for SpHb measurements.
1. Verify that you are using the Intervals Monitoring or Continuous Monitoring profile.
2. Touch the Alarms tab.
3. Touch the SpHb vertical tab.
4. Ensure that the SpHb alarm limit control is set to ON.
Note
If any parameter's alarm limit control is set to OFF, you
cannot adjust alarm limits on the Alarm tab, and no visual or
audio signals will occur for that specific parameter.
5. Enter the desired upper and lower alarm limits for SpHb using the up/down arrow
keys or the keypad.
6. Touch the Home tab.
The new alarm settings take effect immediately.
Set SpHb averaging mode
Touch
in the SpHb frame.
The SpHb frame displays the current mode.
Measure SpHb
WARNING Inaccurate measurement risk. Use only Masimo rainbow SET
sensors and accessories on Masimo-equipped monitors.
WARNING High-intensity lights (including pulsating strobe lights) directed
on the pulse co-oximeter can prevent it from obtaining readings.
WARNING The pulsations from intra-aortic balloon support can increase
the pulse rate displayed on the monitor. Verify the patient's pulse rate
against the ECG heart rate.
WARNING Inaccurate measurement risk. Venous congestion might cause
a low reading of arterial oxygen saturation. Place the sensor on a hand at
heart level to ensure proper venous outflow from a monitored site.
150 Patient monitoring