In the Device Status area, if the icon changes to

and the message

remains, the timer counts down a fixed pause interval of 60 seconds. The alarm

tone returns after the pause interval.

A longer pause interval can be configured in the Advanced tab. You start the

configured pause interval by touching

.

If you responded to an NIBP alarm and multiple NIBP limits have been

exceeded, the alarm tone pauses, the first message clears, and the next NIBP

limit message appears with a countdown timer. A new NIBP alarm tone sounds

after the countdown unless you touch

to dismiss each remaining NIBP

limit message.

2. If multiple alarms are active, a multiple alarm toggle appears in the Device Status

area. The monitor will either autoscroll through the alarm messages, displaying each

for approximately 4 seconds in order of priority, or you can move through the alarms

manually. Respond to multiple alarms manually as follows:

a. Touch

to pause all audio alarms.

The alarm icon changes to

and the timer counts down a fixed pause

interval of 60 seconds. The alarm tone returns after the pause interval.

b. Touch

in the Device Status area to view each alarm in the stack.

Note

The multiple alarm toggle button displays the number of
active alarms inside the alarm icon. A set of dots
indicating the display order of alarms from highest (left)
to lowest (right) priority (as well as the most recent in
the case of multiple alarms of the same priority) appears
below it.

Directions for use

Alarms 87