CO2 specifications (for additional information, refer to the directions for use that accompanies

each accessory)

Sampled gas return/disposal

Sampled gases are not returned to the breathing circuit.

Exhaled gas is exhausted from the monitor's exhaust port.

Dispose of sampled gases according to facility requirements

or local regulations.

Calibration interval

Initial: After 1200 operating hours
Subsequent: After 4000 operating hours or annually

(whichever comes first)

Periodic service

After 30,000 operating hours

etCO2

Units of measure

mmHg, kPa; user-selectable

Display range

0 to 150 mmHg (0.0 to 20.0 kPa)

Resolution

1 mmHg, 0.1 kPa

FiCO2

Units of measure

mmHg, kPa; user-selectable

Display range

0 to 150 mmHg (0.0 to 20.0 kPa)

Resolution

1 mmHg, 0.1 kPa

IPI

Display range

1 to 10

RR

Unit of measure

breaths per minute

Display range

0 to 150 breaths per minute

Accuracy

0 to 70 breaths per minute: ±1 breath per minute
71 to 120 breaths per minute: ±2 breaths per minute
121 to 150 breaths per minute: ±3 breaths per minute

Resolution

1 breath per minute

1

The accuracy specification is maintained to within 4% of the values indicated in the presence of interfering gases in

accordance with ISO 21647, clauses 51.101.3 and 101.1.

2

The accuracy specification is maintained to within 4% of the values indicated in the presence of up to 80% helium

with up to 15% oxygen when tested in accordance with ISO 21647.

3

The accuracy specification is maintained when tested for drift in accordance with ISO 21647, clause 51.101.2.