Welch Allyn 6000 Series Connex Vital Signs Monitor - User Manual User Manual

Page 205

Advertising
background image

Nellcor specifications (for additional information, refer to the directions for use that

accompanies each accessory)

SpO2

Unit of measure

%

Measurement range

1% to 100%

Nellcor sensor accuracy guide

SpO2 measurement accuracy can only be evaluated in vivo

by comparing pulse oximeter readings with SaO2

measurements obtained from simultaneously sampled

arterial blood made using a laboratory CO-oximeter. SpO2

accuracy was validated through breathe-down-equivalent

testing by Covidien using electronic measurements to prove

equivalence to the Nellcor N600x predicate device. The

Nellcor N600x predicate device was validated by performing

human-subject, “breathe-down” clinical trials.

Accuracy

Note

Saturation accuracy varies by sensor type.
The worst case accuracy range of the
module or the attached sensor will apply.

Saturation accuracy (module)

±3 digits
70% to 100%
Adult, Pediatrics: ± 2 digits
Neonate: ± 3 digits
Low Perfusion: 0.02 % to 20 % ± 2 digits

Saturation accuracy (sensors)

Sensor

Accuracy

60% to 80%

MAX-AI, MAX-

PI, MAX-II

±3 digits

70% to 100%

DS-100A

±3 digits

D-YS

Infants, Pediatrics, Adults: ±3 digits
Neonates: ±4 digits

D-YSE

±4 digits

D-YSPD

±4 digits

MAX-AI, MAX-

PI, MAX-II

±2 digits

OXI-A/N

Adults: ±3 digits
Neonates: ±4 digits

OXI-P/I

±3 digits

Pulse rate

Unit of measure

beats per minute

Directions for use

Specifications 199