EarlySense specifications
Operating current
Main power supply:
Minimum: 4.9V
Typical: 5.0V
RTC backup battery voltage:
Minimum: 2.5V
Typical: 3.0V
Maximum: 3.2V
Power consumption at 5V operating current Minimum: 4mA
Maximum: 400mA
Patient movement
Movement during defined period (1.5
minutes)
0 = 0%
L = Up to 40%
M = 40% to 60%
H = 60% to 80%
EH = 80% to 100%
Accuracy
Adult:
0 = 100%, L = 100%, M = 81%, H = 100%, EH = 96%
Pediatric:
0 = 100%, L = 100%, M = 81%, H = 86%, EH = 94%
Averaging period
15 seconds
Respiration rate
Unit of measure
breaths per minute
Measurement range
6 to 45 breaths per minute
1
Accuracy
2
±4% or ±1.5 breaths per minute, whichever is greater
Averaging period
1 minute
Pulse rate
Unit of measure
beats per minute
Measurement range
30 to 170 beats per minute
1
Accuracy
2
±4% or ±5 beats per minute, whichever is greater
Averaging period
1 minute
1
The system detects pulse rate that is > 1.8 times the respiration rate.
2
Total system accuracy including undetected signals equals 90%.