EarlySense specifications

Operating current

Main power supply:
Minimum: 4.9V
Typical: 5.0V
RTC backup battery voltage:
Minimum: 2.5V
Typical: 3.0V
Maximum: 3.2V

Power consumption at 5V operating current Minimum: 4mA

Maximum: 400mA

Patient movement

Movement during defined period (1.5

minutes)

0 = 0%
L = Up to 40%
M = 40% to 60%
H = 60% to 80%
EH = 80% to 100%

Accuracy

Adult:
0 = 100%, L = 100%, M = 81%, H = 100%, EH = 96%
Pediatric:
0 = 100%, L = 100%, M = 81%, H = 86%, EH = 94%

Averaging period

15 seconds

Respiration rate

Unit of measure

breaths per minute

Measurement range

6 to 45 breaths per minute

1

Accuracy

2

±4% or ±1.5 breaths per minute, whichever is greater

Averaging period

1 minute

Pulse rate

Unit of measure

beats per minute

Measurement range

30 to 170 beats per minute

1

Accuracy

2

±4% or ±5 beats per minute, whichever is greater

Averaging period

1 minute

1

The system detects pulse rate that is > 1.8 times the respiration rate.

2

Total system accuracy including undetected signals equals 90%.