Alarm limits

Upper limit range of entry

Lower limit range of entry

Systolic

Adult: 30 to 258 mmHg (4.0 to 34.4 kPa). Factory

default: 220 mmHg (29.3 kPa).
Pediatric: 32 to 160 mmHg (4.3 to 21.3 kPa).

Factory default: 145 mmHg (19.3 kPa).
Neonate: 27 to 120 mmHg (3.6 to 16.0 kPa).

Factory default: 100 mmHg (13.3 kPa).

Adult: 28 to 256 mmHg (3.7 to 34.1 kPa). Factory

default: 75 mmHg (10.0 kPa).
Pediatric: 30 to 158 mmHg (4.0 to 21.1 kPa).

Factory default: 75 mmHg (10.0 kPa).
Neonate: 25 to 118 mmHg (3.3 to 15.7 kPa).

Factory default: 50 mmHg (6.7 kPa).

Diastolic

Adult: 22 to 235 mmHg (2.9 to 31.3 kPa). Factory

default: 110 mmHg (14.7 kPa).
Pediatric: 17 to 130 mmHg (2.3 to 17.3 kPa).

Factory default: 100 mmHg (13.3 kPa).
Neonate: 12 to 105 mmHg (1.6 to 14.0 kPa).

Factory default: 70 mmHg (9.3 kPa).

Adult: 20 to 233 mmHg (2.7 to 31.1 kPa). Factory

default: 35 mmHg (4.7 kPa).
Pediatric: 15 to 128 mmHg (2.0 to 17.1 kPa).

Factory default: 35 mmHg (4.7 kPa).
Neonate: 10 to 103 mmHg (1.3 to 13.7 kPa).

Factory default: 30 mmHg (4.0 kPa).

MAP

Adult: 22 to 255 mmHg (2.9 to 34.0 kPa). Factory

default: 120 mmHg (16.0 kPa).
Pediatric: 17 to 140 mmHg (2.3 to 18.7 kPa).

Factory default: 110 mmHg (14.7 kPa).
Neonate: 12 to 110 mmHg (1.6 to 14.7 kPa).

Factory default: 80 mmHg (10.7 kPa).

Adult: 20 to 253 mmHg (2.7 to 33.7 kPa). Factory

default: 50 mmHg (6.7 kPa).
Pediatric: 15 to 138 mmHg (2.0 to 18.4 kPa).

Factory default: 50 mmHg (6.7 kPa).
Neonate: 10 to 108 mmHg (1.3 to 14.4 kPa).

Factory default: 35 mmHg (4.7 kPa).

SpO2

Adult, Pediatric, Neonate: 52% to 100%. Factory

default: 100%.

Adult, Pediatric, Neonate: 50% to 98%. Factory

default: 90%.

SpHb

Adult, Pediatric, Neonate: 1.5 to 24.5 g/dL (1.5

to 15.0 mmol/L). Factory default: 17.0 g/dL (11.0

mmol/L).

Adult, Pediatric, Neonate: 1.0 to 24.0 g/dL (1.0

to 14.5 mmol/L). Factory default: 7.0 g/dL (4.0

mmol/L).

Pulse rate

Adult: 27 to 300 beats per minute. Factory

default: 120 beats per minute.
Pediatric: 27 to 300 beats per minute. Factory

default: 150 beats per minute.
Neonate: 27 to 300 beats per minute. Factory

default: 200 beats per minute.

Adult: 25 to 298 beats per minute. Factory

default: 50 beats per minute.
Pediatric: 25 to 298 beats per minute. Factory

default: 50 beats per minute.
Neonate: 25 to 298 beats per minute. Factory

default: 100 beats per minute.

Temperature

Adult, Pediatric, Neonate: 87.1 to 110.0°F (30.6

to 43.3°C). Factory default: 101.0°F (38.3°C).

Adult, Pediatric, Neonate: 85.1 to 108.0°F (29.5

to 42.2°C). Factory default: 94.0°F (34.4°C).

etCO2

Adult: 2 to 150 mmHg (0.3 to 20.0 kPa). Factory

default: 60 mmHg (8.0 kPa).
Pediatric: 2 to 150 mmHg (0.3 to 20.0 kPa).

Factory default: 60 mmHg (8.0 kPa).
Neonate: 2 to 150 mmHg (0.3 to 20.0 kPa).

Factory default: 50 mmHg (6.7 kPa).

Adult: 0 to 148 mmHg (0.0 to 19.7 kPa). Factory

default: 15 mmHg (2.0 kPa).
Pediatric: 0 to 148 mmHg (0.0 to 19.7 kPa).

Factory default: 15 mmHg (2.0 kPa).
Neonate: 0 to 148 mmHg (0.0 to 19.7 kPa).

Factory default: 20 mmHg (2.7 kPa).

FiCO2

Adult: 0 to 150 mmHg (0.0 to 20.0 kPa). Factory

default: 8 mmHg (1.1 kPa).
Pediatric: 0 to 150 mmHg (0.0 to 20.0 kPa).

Factory default: 8 mmHg (1.1 kPa).
Neonate: 0 to 150 mmHg (0.0 to 20.0 kPa).

Factory default: 5 mmHg (0.7 kPa).

None