relay. If you select
Low
, high-, medium-, and low-
priority alarms activate a nurse call relay.
Audio
(vertical tab)
Allow user to turn off general audio
Select to allow clinicians to turn off all audio notification
for alarms. This control is on the Alarms tab (on the
General tab).
Minimum alarm volume
Select the minimum alarm volume available. If you select
High
, then
Medium
and
Low
are not available to
the clinician.
Audio pause time
Specify the amount of pause time that is added to the
60-second pause time. When a clinician pauses an audio
alarm tone, the tone is paused for the combined amount
of time.
Delays
(vertical tab)
Note Alarm delays are available only for the parameters
supported by your configuration.
SpO2 alarm condition delay
Specify the minimum amount of time that an SpO2 alarm
condition must be active before audio and visual signals
occur.
SatSeconds is available with Nellcor SpO2 sensors. If
you select Off or 10 seconds, SatSeconds is disabled,
and it is removed from the SpO2 tab in the Alarms tab.
SpO2 pulse rate alarm condition delay
Specify the minimum amount of time that a Pulse rate
alarm condition measured by the SpO2 sensor must be
active before audio and visual signals occur.
Motion pulse rate alarm condition delay
Specify the minimum amount of time that a Pulse rate
alarm condition measured by the patient movement
sensor must be active before audio and visual signals
occur.
SpHb alarm condition delay
Specify the minimum amount of time that an SpHb alarm
condition must be active before audio and visual signals
occur.
Motion respiration alarm condition delay
Specify the minimum amount of time that a respiration
alarm condition must be active before audio and visual
signals occur (EarlySense configuration).
Motion pulse rate low confidence alarm delay
Specify the minimum amount of time that a pulse rate
low confidence alarm condition must be active before
audio and visual signals occur (EarlySense
configuration).
Motion respiration low confidence alarm delay
Specify the minimum amount of time that a respiration
low confidence alarm condition must be active before
audio and visual signals occur (EarlySense
configuration).
etCO2 alarm condition delay
Specify the minimum amount of time that an etCO2
alarm condition must be active before audio and visual
signals occur.
Respiration alarm condition delay
Specify the minimum amount of time that an RR alarm
condition must be active before audio and visual signals
occur.
No breath detected alarm delay
If configured with a CO2 module, specify the amount of
time for the device to wait after receiving a "valid
breath" message before activating the "No breath
detected" physiological alarm.
If configured for RRa, specify the time that the Masimo
module will wait before sending a respiration pause
Directions for use