Message

Possible cause

Suggested action

The bed sensor is upside down.

The bed sensor was not placed properly under

the mattress, mattress cover, or mattress pad

Turn the sensor right-side up.

Cannot measure patient. Lost or

unstable signal.

The sensor is not under the patient's chest

(Low confidence condition)

Reposition the sensor under the

patient's chest.

The sensor is turned 90 degrees, placed

vertically (Low confidence condition)

Place the sensor horizontally

under the patient's mattress,

with the cable extending toward

the head of the bed.

There is excessive patient movement (Low

confidence condition)

Check the patient.
Replace the bed sensor.

An unqualified mattress type is in use

Switch to a qualified mattress

type.

Cannot measure respiration.

The sensor is not under the patient's chest

(Low confidence condition)

Adjust the position of the sensor

under the patient's chest.

The sensor is turned 90 degrees, placed

vertically (Low confidence condition)

Place the sensor horizontally

under the patient's mattress,

with the cable extending toward

the head of the bed.

There is excessive patient movement (Low

confidence condition)

Check the patient.
Replace the bed sensor.
Use an alternative sensor to

monitor respiration.

An unqualified mattress type is in use

Switch to a qualified mattress

type.

Cannot measure pulse rate.

The sensor is not under the patient's chest

(Low confidence condition)

Adjust the position of the sensor

under the patient's chest.

The sensor is turned 90 degrees, placed

vertically (Low confidence condition)

Place the sensor horizontally

under the patient's mattress,

with the cable extending toward

the head of the bed.

There is excessive patient movement (Low

confidence condition)

Check the patient.
Replace the bed sensor.
Use an alternative sensor to

monitor pulse rate.

An unqualified mattress type is in use

Switch to a qualified mattress

type.

EarlySense not functional.

A module error occurred

Call for service.

Bed sensor expires in ...

The bed sensor expires soon

Replace the bed sensor before it

expires.

The sensor has expired.

The bed sensor has expired

Replace the sensor.