CO2 specifications (for additional information, refer to the directions for use that accompanies

each accessory)

Accuracy

0 to 70 breaths per minute: ±1 breath per minute
71 to 120 breaths per minute: ±2 breaths per minute
121 to 150 breaths per minute: ±3 breaths per minute

Resolution

1 breath per minute

1

The accuracy specification is maintained to within 4% of the values indicated in the presence of interfering gases in

accordance with ISO 21647, clauses 51.101.3 and 101.1.

2

The accuracy specification is maintained to within 4% of the values indicated in the presence of up to 80% helium

with up to 15% oxygen when tested in accordance with ISO 21647.

3

The accuracy specification is maintained when tested for drift in accordance with ISO 21647, clause 51.101.2.

EarlySense specifications

Sensor

Dimensions

300mm x 210mm x 2.5mm

Weight

160 g

Material

Polycarbonate + ABS

Connector

RS232 (S11M03-P04M500-5260(ODU))

Sensor power supply

Electrical absolute maximum current, DC in Minimum: –0.3V

Maximum: 5.5V

Operating current

Minimum: 4.9V
Typical: 5.0V
Maximum: 5.2V

Power consumption at 5V operating current Minimum: 1mA

Maximum: 4mA

Module

Dimensions

135mm x 75mm x 10mm

Weight

800 g

Host connection

USB Mini-B

Module power supply

Electrical absolute maximum current, DC in Minimum: –0.3V

Maximum: 5.5V

214 Specifications