NIBP specifications

Pediatric: 280 mmHg (37.3 kPa) (StepBP, SureBP)
Neonate: 130 mmHg (17.3 kPa) (StepBP)

Blood pressure determination time

Typical: 15 seconds
Maximum: 150 seconds

Blood pressure accuracy

Meets or exceeds ANSI/AAMI SP10:2002 standards for noninvasive blood

pressure accuracy, ±5 mmHg (0.7 kPa) mean error, 8 mmHg (1.1 kPa)

standard deviation

Mean Arterial Pressure (MAP) range
The formula used to calculate MAP yields

an approximate value.

Adult: 23 to 230 mmHg (3.1 to 30.7 kPa) (StepBP, SureBP)
Pediatric: 23 to 230 mmHg (3.1 to 30.7 kPa) (StepBP, SureBP)
Neonate: 13 to 110 mmHg (1.7 to 14.7 kPa) (StepBP)

Pulse rate range (using blood pressure

determination)

Adult: 30 to 200 beats per minute (StepBP, SureBP)
Pediatric: 30 to 200 beats per minute (StepBP, SureBP)
Neonate: 35 to 220 beats per minute (StepBP)

Pulse rate accuracy (using blood pressure

determination)

±5.0% (±3 beats per minute)

Overpressure cutoff

Adult: 300 mmHg ±15 mmHg (40.0 kPa ±2.0 kPa)
Pediatric: 300 mmHg ±15 mmHg (40.0 kPa ±2.0 kPa)
Neonate: 150 mmHg (20.0 kPa) maximum

SureTemp Plus temperature module specifications

Units of measure

°F, °C; user-selectable

Temperature range

80°F to 110°F (26.7°C to 43.3°C)

Calibration accuracy

±0.2°F (±0.1°C) (Direct mode)

Braun ThermoScan PRO 4000 thermometer specifications (refer to manufacturer's directions for

use for additional information)

Units of measure

°F, °C; user-selectable

Temperature range

68°F to 108°F (20°C to 42.2°C)

Calibration accuracy

±0.4°F (±0.2°C) for temperatures ranging from 95.9°F to 107.6°F
(35.5°C to 42°C)

±0.5°F (±0.3°C) for temperatures outside of this range

Display resolution

0.1°F or °C

198 Specifications