Welch Allyn 6000 Series Connex Vital Signs Monitor - User Manual User Manual
Page 165
2. Locate the battery cover, indicated by
.
3. Insert a coin into the slot and push to open. Choose a coin that fits comfortably in
the slot.
4. Pull the battery out by pulling the battery label, which is visible when you open the
battery cover.
5. Slide in the new battery. Ensure that you insert the new battery in the same
orientation as the old battery.
6. Replace the battery cover by positioning the end below
and then pressing
firmly on the opposite end.
Note
New batteries are only 30 percent charged. Therefore, connect the monitor
to AC power immediately after inserting a new battery.
WARNING Risk of fire, explosion, and burns. Do not short-circuit, crush,
incinerate, or disassemble the battery pack. Never dispose of batteries in
refuse containers. Always recycle batteries according to local regulations.