No. Feature

Description

Green: The battery is charged.

Amber: The battery is charging.

2

Ethernet RJ-45

Provides a hardwired connection to the computer network.

3

USB client

Provides a connection to an external computer for testing and

software upgrades.

4

Nurse call

Provides a connection to the hospital nurse call system. (Not

available on the 6300 model.)

5

Fan exhaust

6

Ground lug (equipotential terminal)

Provided for electrical safety testing and as a means for

connection of a potential-equalization conductor.

7

Power connection

Provides an external AC power connection.

8

Mobile stand mounting hardware

Secures the mounting plate to the monitor.

9

Recess for mounting plate

Secures the monitor when the monitor is mounted on the mobile

stand or wall.

10

USB connector door

Provides access to host USB connections for optional

accessories.

11

Fan intake

12

Speaker

Provides tones. A piezo beeper inside the monitor provides

backup.

22 Physical design

Welch Allyn Connex

®

Vital Signs Monitor 6000 Series™