Using multiple interfaces, 1 using multiple interfaces – Bronkhorst EL-FLOW Prestige User Manual

Page 15

Advertising
background image

Bronkhorst High-Tech B.V.

EL-FLOW Prestige

15

9.17.084

Via the 9-pin sub-D side connector the instrument can be operated by means of:

Analog interface (section 3.3): 0…5 Vdc; 0…10 Vdc; 0…20 mA or 4…20 mA
Digital RS232 interface (section 3.4 and document 9.17.027): (FLOW-BUS (Propar) protocol)
Digital RS485 interface (section 3.5): (Modbus RTU, Modbus ASCII or FLOW-BUS protocols)

The following optional field bus interfaces can be installed:

FLOW-BUS interface (section 3.5 and document 9.17.024)
Modbus (RTU or ASCII) interface (section 3.5 and document 9.17.035)
PROFIBUS DP interface (section 3.6 and document 9.17.025)
DeviceNet

TM

interface (section 3.6 and document 9.17.026)

EtherCAT® interface (section 3.6 and document 9.17.063)
PROFINET interface (section 3.6 and document 9.17.095)

The interpretation of the LED indications and use of the micro switch button on top of the instrument is discussed in section 3.7

and section 3.8 respectively.

3.2.1

Using multiple interfaces

The analog interface is always present on EL-FLOW Prestige instruments. An interface to any available field bus is optional.

Operation via analog interface, RS232/RS485 (side connector) and an optional field bus (top connector) can be performed at the

same time. When using multiple interfaces, reading of parameters can be done simultaneously. When changing a parameter value,

the last value sent by any of the interfaces will be valid.

Control mode

A controller setpoint is accepted from either the analog or digital interface, but not both. Analog or digital operation is selected at

ordering and indicated on the instrument backside label. The inactive setpoint source is indicated between brackets (…), see

example below. The parameter ‘Control Mode’ indicates from which source a controller setpoint is accepted: analog or digital. See

section 4.2.2 for more information regarding the ‘Control Mode’ parameter.

Factory communication settings

The factory selected communication and side connector pinning settings are indicated on the instrument backside label. See

example below for a description of the communication information indicated on the instrument label:

1. Model key (see explanation of the model key in section 1.5)

2. Field bus (top connector) (example: none)

3. Customized IO settings (pin 5), see section 4.4.1 for more

information (example: default setting)

4. Analog interface (pin 2, 3) (example: inactive setpoint

source, measure always available)

5. Digital interface (pin 1, 6) (example: RS232, active setpoint

source)

6. Side connector digital interface settings (protocol, medium,

baud rate, parity) (example: FLOW-BUS (Propar)

communication over RS232 interface with baud rate 115200

Baud and no parity)

Advertising