Bronkhorst EL-FLOW Prestige User Manual

Page 36

Advertising
background image

Bronkhorst High-Tech B.V.

EL-FLOW Prestige

9.17.084

36

Capacity

Type

Access

Range

FlowDDE

FLOW-BUS

Modbus

Float

RW

1E-10...1E+10

21

1/13

0x8168...0x8169/33129...33130

This parameter sets the maximum readout/control value (100%) for the active ‘

24

Fluid Number’ in readout units corresponding to

the '

129

Capacity Unit' parameter. The '

21

Capacity' is scaled when '

178

Pressure Inlet', '

181

Fluid Temperature' or '

25

Fluid Name' (or any

of the mixture parameter) are changed for the active fluidset.

Capacity Unit

Type

Access

Range

FlowDDE

FLOW-BUS

Modbus

Unsigned char[7]

RW

See below

129

1/31

0x81F8...0x81FB/33273...33276

For EL-FLOW Prestige the following units can be set:

Mass flow

Normal volume flow

(1.01325 bar(a), 0 °C)

Standard volume flow

(1.01325 bar(a), 20 °C)

(Custom) volume flow

(

246

Capacity Unit Pressure,

245

Capacity Unit Type

Temperature)

ug/h, ug/min, ug/s,

mg/h, mg/min, mg/s,

g/h, g/min, g/s,

kg/h, kg/min, kg/s

uln/h, uln/min, uln/s,

mln/h, mln/min, mln/s,

ln/h, ln/min, ln/s,

ccn/h, ccn/min, ccn/s,

mm3n/h, mm3n/m, mm3n/s,

cm3n/h, cm3n/m, cm3n/s,

m3n/h, m3n/min, m3n/s,

scfh, scfm, scfs,

sccm, slm

uls/h, uls/min, uls/s,

mls/h, mls/min, mls/s,

ls/h, ls/min, ls/s,

ccs/h, ccs/min, ccs/s,

mm3s/h, mm3s/m, mm3s/s,

cm3s/h, cm3s/m, cm3s/s,

m3s/h, m3s/min, m3s/s

ul/h, ul/min, ul/s,

ml/h, ml/min, ml/s,

l/h, l/min, l/s,

cc/h, cc/min, cc/s,

mm3/h, mm3/m, mm3/s,

cm3/h, cm3/m, cm3/s,

m3/h, m3/min, m3/s,

cfh, cfm, cfs

Unit type list

Due to compatibility the maximum string length is limited to 7 characters. Therefore some unit names are

truncated. For instance mm3n/m means mm

3

n

/min.

Controller Speed (Kspeed)

Type

Access

Range

FlowDDE

FLOW-BUS

Modbus

Float

RW

0.2...5

254

114/30

0xF2F0...0xF2F1/62193...62194

This parameter sets the controller speed factor for the selected fluidset. The '

254

Kspeed' parameter is set from factory between

value '0.5' (slow) and '2' (fast). The default value is '1'. Slower or faster settings are possible between values 0.2 and 5 at the

customer's responsibility.

Actual Temperature

Type

Access

Range

FlowDDE

FLOW-BUS

Modbus

Float

R

-250...500

142

33/7

0xA138...0xA139/41273...41274

Indication of the actual ambient temperature in °C.

Pressure

Type

Access

Range

FlowDDE

FLOW-BUS

Modbus

Float

RW

0...3.4E+38

143

33/8

0xA140...0xA141/41281...41282

Indication of the actual (inlet) pressure in bar(a). By default this parameter value is equal to the value of '

178

Inlet pressure'. This

parameter can be used for active pressure correction by setting parameter '

354

Conversion Condition Selection' to value 3 and

writing the actual pressure to '

143

Pressure'. After power reset the parameter value is reset to the value of '

178

Inlet pressure'.