Advanced parameters and properties, Advanced measurement and control parameters, Special instrument parameters – Bronkhorst EL-FLOW Prestige User Manual

Page 33: 2 advanced parameters and properties, 1 advanced measurement and control parameters, 2 special instrument parameters

Advertising
background image

Bronkhorst High-Tech B.V.

EL-FLOW Prestige

33

9.17.084

4.2

Advanced parameters and properties

4.2.1

Advanced measurement and control parameters

Measured Value (Fmeasure)

Type

Access

Range

FlowDDE

FLOW-BUS

Modbus

Float

R

-3.4E+38…

3.4E+38

205

33/0

0xA100…0xA101/ 41217…41218

Floating point variable of the ‘

8

Measured Value’. The ‘

205

Fmeasure’ parameter shows the measured value in the capacity and

capacity unit for which the instrument has been set. The ‘

205

Fmeasure’ parameter is dependent of ‘

129

Capacity Unit’ and ‘

22

Sensor

Type’.

Setpoint (Fsetpoint)

Type

Access

Range

FlowDDE

FLOW-BUS

Modbus

Float

RW

0…3.4E+38

206

33/1

0xA119…0xA11A/ 41241…41242

Floating point variable of the ‘

9

Setpoint’. The ‘

206

Fsetpoint’ parameter shows the setpoint in the capacity and capacity unit for

which the instrument has been set. The ‘

206

Fsetpoint’ parameter is dependent of ‘

129

Capacity Unit’ and ‘

22

Sensor Type’.

Valve Output

Type

Access

Range

FlowDDE

FLOW-BUS

Modbus

Unsigned long

RW

0…

16777215

55

114/1

0x001F/32

This parameter is the digital steering signal for driving the control valve, where 0…16777215 corresponds with 0…100%.

Analog Input

Type

Access

Range

FlowDDE

FLOW-BUS

Modbus

Unsigned int

R

0…65535

11

1/3

0x0023/36

Depending on the analog mode, 0…5 Vdc, 0…10 Vdc, 0…20 mA or 4…20 mA is converted to a number in the range 0…32000. The

digitized ‘

11

Analog Input’ is in the same range as ‘

8

Measured Value’ (0…32000 corresponds to 0…100%). This parameter can be used

as setpoint or slave factor when the instrument is used as ‘analog slave’ or ‘FLOW-BUS analog slave’, see ‘

12

Control Mode’.

4.2.2

Special instrument parameters

All parameters described in this section have influence on the behavior of the EL-FLOW Prestige. Please be aware

that wrong settings can disorder the output. To avoid unintentional changes, some parameters are locked (shown by

the symbol). To unlock parameters set parameter ‘

7

Init Reset’ to ‘Unlocked’.

Init Reset

Type

Access

Range

FlowDDE

FLOW-BUS

Modbus

Unsigned char

RW

82/64

7

0/10

0x000A/11

The ‘

7

Init Reset’ parameter is used to unlock secured parameters for writing (see the symbol). This parameter can be set to the

following values:

Value 82: Locked mode, secured parameters are read-only
Value 64: Unlocked mode, secured parameters are writeable and readable

This parameter is always set to ‘Locked’ (value 82) at power-up.