Troubleshooting and service, Diagnostics, 5troubleshooting and service – Bronkhorst EL-FLOW Prestige User Manual

Page 46: 1 diagnostics

Advertising
background image

Bronkhorst High-Tech B.V.

EL-FLOW Prestige

9.17.084

46

4. Write the new value to parameter ‘

86

IOStatus’

5. Set parameter ‘

7

Init Reset’ to 82

Note: do not use this procedure if customized IO options -C5S- (digital output, enabled by setpoint), -I3C- (digital input, controller

mode valve close) or - I8C- (digital input, controller mode valve purge) are installed.

5

Troubleshooting and service

5.1

Diagnostics

When errors or warnings occur, connect the instrument to the Bronkhorst support software to determine the cause

of the error or warning. For any fluidset related errors/warnings, make connection to the FlowTune software for

diagnostics. For other errors/warnings, connect the instrument to FlowPlot and open the 'Alarm & Count' section

under 'Instrument Settings' to determine the cause of the alarm/warning message.

Firmware Version

Type

Access

Range

FlowDDE

FLOW-BUS

Modbus

Unsigned char[6]

R

-

105

113/5

0xF128…0xF12A/ 61737…61739

Revision number of firmware.

Identification Number

Type

Access

Range

FlowDDE

FLOW-BUS

Modbus

Unsigned char

RW

0...255

175

113/12

0x0E2C/3629

Bronkhorst (digital) device/instrument identification number. For EL-FLOW Prestige instruments the following values are

applicable:

Value 7: DMFC (Digital Mass Flow Controller)
Value 8: DMFM (Digital Mass Flow Meter)

Device Type

Type

Access

Range

FlowDDE

FLOW-BUS

Modbus

Unsigned char[6]

R

-

90

113/1

0xF108…0xF10A/ 61705…61707

The device type: DMFC or DMFM is stored in this read-only parameter.