Bronkhorst High-Tech B.V.

EL-FLOW Prestige

43

9.17.084

Side connector RS485 communication

Change the baud rate, node address or parity of the side-connector RS485 (FLOW-BUS or Modbus) interface (if installed) with the

following parameters in the 'Configuration Mode'.

Note: an instrument with 9-pin sub-D side connector set for RS485 FLOW-BUS or Modbus communication will not

respond when connecting to an RS232 configuration. If the instrument is not set for RS232 communication, use the

micro switch on top of the instrument to overrule the custom settings and switch to RS232 communication settings:

press and hold the micro switch at power-up and wait (12…16 sec) until both green and red LEDs flash (0.2 sec

on, 0.2 sec off). Release the switch to activate the ‘Configuration Mode’. In the ‘Configuration Mode’ the bus type

and baud rate for the 9-pin sub-D side connector are set to RS232 FLOW-BUS (Propar) at 38400 Baud. The

‘Configuration Mode’ remains active after power down. Use the same procedure to deactivate the ‘Configuration

Mode’.

Fieldbus 2 address

Type

Access

Range

FlowDDE

FLOW-BUS

Modbus

Unsigned char

RW

0…255

309

124/10

0xFC50/64593

For the accepted values see the applicable field bus in chapter 3.

Fieldbus 2 baudrate

Type

Access

Range

FlowDDE

FLOW

-BUS

Modbus

Unsigned long

RW

0…10000000000

310

124/9

0xFC48...0xFC49/64585...64586

For the accepted values see the applicable field bus in chapter 3.

Fieldbus 2 parity

Type

Access

Range

FlowDDE

FLOW-BUS

Modbus

Unsigned char

RW

0…2

336

124/12

0xFC60/64609

The following values are accepted:

0 – no parity

1 – odd parity

2 – even parity

4.4

Special instrument features

4.4.1

Customized IO options (pin 5)

EL-FLOW Prestige instruments offer various customized input/output functions through pin 5 of the 9-pin sub-D side connector as

an option. A number of the (factory installed) programmable IO options are offered as standard, these options are described in this

section. The last three characters of the model key (presented on the back-side label of the instrument) indicate the installed IO

configuration, see also section 1.5. The standard options are described below (the default selection is …-A1V). Refer to document

9.16.118 for the applicable hook-up diagrams.

The customized IO options are factory installed, as indicated by last three characters of the model key (presented on

the back-side label of the instrument). These IO options cannot be changed manually.

A1V

0…10 Vdc output, controller (default selection)

Analog signal for pump or external valve steering (control signal only).

When the controller output is used for pump or external valve steering (Mass Flow Meters only), make sure the

231

Valve Maximum’ is set to 0.3 [A]. For Mass Flow Controllers, the controller output is limited to a value below

10 Vdc due to the maximum valve current restriction.

B1V

4…20 mA output, controller

Analog signal for pump or external valve steering (control signal only).