Troubleshooting, 2 troubleshooting – Bronkhorst EL-FLOW Prestige User Manual

Page 47

Advertising
background image

Bronkhorst High-Tech B.V.

EL-FLOW Prestige

47

9.17.084

5.2

Troubleshooting

In this section a number of possible errors/malfunctions of the instrument are listed.

LED indications

See section 3.7 for interpretation of specific LED indications of the instrument

Factory reset

In case (re)setting of the instrument has led to unexpected or non-recoverable behavior, it is possible to reset the instrument to the

settings applied at the factory during production. Use the 'Restore Settings' tab under 'Instrument Settings' in FlowPlot or use the

micro switch on top of the instrument (section 3.8) to restore the instrument to the original factory settings. Note that any

changes made to the instrument communication settings will not be restored after a factory reset. If digital communication with

the instrument can not be re-established, see section 3.4 to overrule the actual 9-pin sub-D communication settings with the

'Configuration Mode' (using the micro switch) and use the RS232 communication mode to re-establish communication.

Troubleshooting list

Symptom

Possible cause

Action

No power (LEDs not burning)

No power supply

Check power supply
Check cable connection

Internal fuse blown due to long lasting

shortage

Return to factory

Zero output signal

No setpoint accepted, incorrect control

mode

Check analog/digital control mode

(section 4.2.2)

No setpoint given, setpoint too low

Give setpoint >2%

Valve in 'Safe State' (normally closed

valves)

Remove cause of 'Valve Safe

State' (see section 3.1.2)

Pcb/sensor failure

Return to factory

Laminar flow element blocked/

contaminated

Return to factory

Maximum output signal (131.07%)

Flow too high, valve fully open

Close valve

Pcb/sensor failure

Return to factory

Valve in 'Safe State' (normally opened

valves)

Remove cause of 'Valve Safe

State' (see section 3.1.2)

Setpoint cannot be achieved

Pressure difference across instrument

insufficient to achieve requested flow rate

Check inlet/outlet pressure

Valve blocked or damaged

Return to factory

Stoppage

Remove contamination

Flow signal is gradually decreasing without

setpoint change

Pressure difference across instrument

insufficient to achieve requested flow rate

Check inlet/outlet pressure

Process gas condensation occurs

Decrease pressure or increase gas

temperature

Oscillating controller output

Controller adjustment wrong, PID settings

too fast

Set '

254

Controller Speed (Kspeed)' to

lower value

Inlet/outlet pressure too high or too low

Adjust pressure/set correct

instrument pressure with FlowTune

Wrong gas selected

Set correct process gas with FlowTune

Inlet pressure control is oscillating

Replace pressure regulator or increase

buffer volume between controller

instruments (section 2.3)

'Cross-talk' with other controller

instruments in process line

Increase buffer volume between

controller instruments (section 2.3)

Valve is damaged

Return to factory

Non-zero output signal when valve is closed Valve is leaking (through valve)

Check valve specifications for leak-

Advertising