Bronkhorst High-Tech B.V.

EL-FLOW Prestige

9.17.084

28

3.9.2

Basic measurement and control parameters

The list below provides the most basic parameters for digital communication with the instrument.

The parameters below are most easily accessible via FlowPlot or FlowView software or any of the Bronkhorst

readout/control units (BRIGHT, E-8000).

Measured Value (Measure)

Type

Access

Range

FlowDDE

FLOW-BUS

Modbus

Unsigned int

R

0…41942

(65535*)

8

1/0

0x0020/33

The ‘

8

Measured Value’ indicates the amount of mass flow metered by the instrument. The signal of 0…100% is presented in a range

of 0…32000. The maximum measured value output is 131.07%, which is 41942. A floating point variable of the measured value,

205

Fmeasure’, is also available in the capacity and capacity unit for which the instrument has been set, see section 4.2.1.

*In case the instrument is prepared for bi-directional measurement, the negative signals with an output range of

-73.73...-0.003% are presented in a range of 41943…65535 (so value 65535 represents -0.003%), whereas the

positive signals 0…131.07% are still presented in the range 0…41942. (In FlowDDE the numbers are converted to

negative values automatically).

Setpoint

Type

Access

Range

FlowDDE

FLOW-BUS

Modbus

Unsigned int

RW

0…32000

9

1/1

0x0021/34

The ‘

9

Setpoint’ is used to set the required mass flow rate for the controller. The signals have the same range as the ‘

8

Measured

Value’, only the setpoint is limited between 0 and 100% (0…32000). A floating point variable of the setpoint, ‘

206

Fsetpoint’, is also

available in the capacity and capacity unit for which the instrument has been set, see section 4.2.1.

3.9.3

Basic identification parameters

User Tag

Type

Access

Range

FlowDDE

FLOW-BUS

Modbus

Unsigned char[16]

RW

-

115

113/6

0xF130…0xF136/ 61745…61751

The ‘

115

User Tag’ parameter allows the user to give the instrument a custom tag name, with a maximum of 16 characters.

Customer Model

Type

Access

Range

FlowDDE

FLOW-BUS

Modbus

Unsigned char[16]

RW

-

93

113/4

0xF120…0xF127/ 61729…61736

This parameter is used to add extra information to the model number information, such as a customer-specific model number.

Serial Number

Type

Access

Range

FlowDDE

FLOW-BUS

Modbus

Unsigned char[20]

R

-

92

113/3

0xF118…0xF11F/ 61721…61728

This parameter consists of a max. 20-byte string with instrument serial number for identification, e.g.: ‘M1111111A’.

BHT Model Number

Type

Access

Range

FlowDDE

FLOW-BUS

Modbus

Unsigned char[27]

RW

-

91

113/2

0xF111…0xF117/ 61713…61719

This parameter shows the Bronkhorst instrument model type information.