Starting up, Check functional properties, Check operating conditions – Bronkhorst EL-FLOW Prestige User Manual

Page 9: Rated pressure test inspection, Sealings, Environmental ratings, Piping requirements, 2starting up, 1 check functional properties, 2 check operating conditions

Advertising
background image

Bronkhorst High-Tech B.V.

EL-FLOW Prestige

9

9.17.084

2

Starting up

2.1

Check functional properties

Before installing your EL-FLOW Prestige mass flow meter/controller it is important to

read the label on the backside of the instrument and check:

Flow rate
Fluid to be measured
Up- and downstream pressures
Temperature
Valve type (N.C.= Normally Closed, N.O.= Normally Opened)
Input/output signal (see also section 2.6)

2.2

Check operating conditions

2.2.1

Rated pressure test inspection

Each EL-FLOW Prestige instrument is pressure tested to at least 1.5 times the maximum operating pressure of the instrument. Each

instrument is helium leak tested to at least 2x10

-9

mbar l/s Helium outboard.

The tested pressure is stated on the instrument with a red-colored sticker.

Before installation, make sure that the test pressure is in accordance with

normal safety factors for your application. If there is no Pressure Testing

Sticker on the device or if the test pressure is incorrect, the instrument

should not be mounted in the process line and be returned to the factory.

2.2.2

Sealings

EL-FLOW Prestige instruments are equipped from factory with sealings compatible with the requested gas type. However the

instruments have multi-gas/multi-range functionality on board. If another gas or mixture is used, always make sure that the gas/

mixture is compatible with the installed sealing materials. Refer to section 4.1 for a compatibility list.

2.2.3

Environmental ratings

EL-FLOW Prestige instruments are suitable for use in conditions between -10…70 °C and relative humidity of 10…90 %RH, unless

specified otherwise. Please note that EL-FLOW Prestige instruments have an ingress protection of IP-40, implying that the

electronics housing and electrical connection do not offer any protection against moist environments.

Make sure that process gases do not condensate in the instrument due to (changing) environmental conditions,

condensed process fluids may seriously harm the instrument's functionality! E.g. do not use moist air as process gas

at low temperatures.

2.3

Piping requirements

For reliable measurement always make sure the fluid stream is clean. Use

filters to assure a moisture-, oil- and particle free gas stream. Recommended

pore-size: 5 µm. If back flow can occur, a downstream filter and a check valve

are recommended too. For high flow rates select a suitable filter size, to avoid

too high pressure drop.

Do not install small diameter piping on high flow rates, because the inlet jet flow may affect the accuracy

Do not mount abrupt angles directly on in- and outlet, especially not on high flow rates. At least 10 pipe diameters

Advertising