Advanced operation

4.1

Sealing material compatibility

EL-FLOW Prestige instruments are equipped from factory with sealings compatible with the requested gas type. However the

instruments have multi-gas/multi-range functionality on board, if another gas or mixture is used, always make sure that the gas/

mixture is compatible with the installed sealing materials. As factory standard, the instruments are provided with Viton® static seals

with FFKM dynamic seals. EPDM sealed instruments or full Kalrez®/FFKM are available as an option, see the model key indication

...-V-..., ...-E-... or ...-K-... on the instrument back-side label. Each sealing material has its specific compatibility with used process

gases. See the table below for compatibility with the 25 most used process gases:

Gas

Modelkey: ...-V-...

Static seals: Viton®

Dynamic seals: FFKM

(viton for n.o. valves or

FDA/USP class VI)

Modelkey: ...-E-...

Static seals: EPDM

Dynamic seals: EPDM

Modelkey: ...-K-...

Static seals: Kalrez®/FFKM

Dynamic seals: FFKM

Remarks

Air
Ar (Argon)
CH

4

(Methane)

CO (Carbon monoxide)
CO

2

(Carbon dioxide)

max: 10 bar(g), 50°C

max: 10 bar(g), 50°C

C

2

F

6

(Freon-116)

C

2

H

2

(Acetylene)

/ *

* FFKM dyn. seal only

C

2

H

4

(Ethylene)

max: 10 bar(g)

max: 10 bar(g)

max: 10 bar(g)

C

2

H

6

(Ethane)

C

3

H

6

#2 (Propylene)

max: 10 bar(g)

max: 58 bar(g)

C

3

H

8

(Propane)

Cl

2

(Chlorine)

Dry gas only

He (Helium)
H

2

(Hydrogen)

H

2

S (Hydrogen sulfide)

NF

3

(Nitrogen

trifluoride)
NH

3

(Ammonia)

NO (Nitric oxide)
N

2

(Nitrogen)

N

2

O (Nitrous oxide)

max: 20 bar(g), 50°C

O

2

(Oxygen)

SF

6

(Sulfur

hexafluoride)
SiH

4

(Silane)

/ *

* FFKM dyn. seal only

Xe (Xenon)

Sealing material compatibility list

Do not use process gases/mixtures that are incompatible with the sealing materials the EL-FLOW Prestige

instrument is equipped with and do not exceed the indicated maximum operating pressure/temperature. Using the

instrument outside the given operating limits may lead to serious damage to the instrument.