Bronkhorst High-Tech B.V.

EL-FLOW Prestige

9.17.084

22

3.5.2

Software

When using a pc to communicate with EL-FLOW Prestige instruments only the FLOW-BUS protocol is supported by Bronkhorst

software. When using Modbus operation, software from third parties, such as LabVIEW, ModScan or a Modbus PLC must be used

to serve as Modbus master.

Note: an instrument with 9-pin sub-D side connector set for RS485 FLOW-BUS or Modbus communication will not

respond when connecting to an RS232 configuration. If the instrument is not set for RS232 communication, use the

micro switch on top of the instrument to overrule the custom settings and switch to RS232 communication settings:

press and hold the micro switch at power-up and wait (12…16 sec) until both green and red LEDs flash (0.2 sec

on, 0.2 sec off). Release the switch to activate the ‘Configuration Mode’. In the ‘Configuration Mode’ the bus type

and baud rate for the 9-pin sub-D side connector are set to RS232 FLOW-BUS (Propar) at 38400 Baud. The

‘Configuration Mode’ remains active after power down. Use the same procedure to deactivate the ‘Configuration

Mode’.

3.5.3

Baud rate, node address and parity setup

EL-FLOW Prestige instruments are configured from factory as indicated on the instrument back-side label. If there is a need of

changing any of the specified RS485 settings, see the tables below for the supported configurations. The default selections are

presented in bold.

Mode:

Digital

Interface/medium:

RS485

Protocol:

FLOW-BUS

Modbus RTU

Modbus ASCII

Baud rate:

187500

400000

9600

19200

38400

56000

57600

115200

128000

256000

9600

19200

38400

56000

57600

115200

128000

256000

Node address:

3…125

1…247

1…247

Parity:

None

None; Even; Odd

None; Even; Odd

RS485 FLOW-BUS/Modbus communication options

Changing RS485 settings of the RJ-45 top connector interface

In case the FLOW-BUS or Modbus RJ-45 field bus connector is used for bus communication,

the node address can be easily set by using the rotary switches on the side of the

instrument. Use the ‘MSD’ (Most Significant Digit) to set the ‘tens’ of the bus-address and

the ‘LSD’ (least Significant Digit) to set the ‘unit’ of the bus-address (the example on the right

reads ‘63’). Set the rotary switches to '00' for automatic installation. Refer to the

corresponding field bus manual, document 9.17.024 (FLOW-BUS) or document 9.17.035

(Modbus) for more details.

For changing the baud rate or parity settings use the RS232 interface to change the

corresponding parameters (see section 4.3.2).

Changing RS485 settings of the 9-pin sub-D side connector interface

In case the 9-pin sub-D side connector is set for RS485 communication, the baud rate or node address can be changed by using the

micro switch or by changing the settings in the ‘Configuration Mode’. Refer to section 3.8 for changing node address and baud rate

with the micro switch. Other communication parameters can be changed only in the ‘Configuration Mode’. Activate the

‘Configuration Mode’ by pressing the micro switch at start-up according the description in section 3.5.2 above. In ‘Configuration

Mode’ the bus type and baud rate are set to RS232 FLOW-BUS (Propar) at 38400 Baud. Change the appropriate parameters as

described in section 4.3.2. When finished, deactivate the ‘Configuration Mode’ using the same procedure. Now the instrument is

ready to use in the desired configuration with the adjusted baud rate, node address or parity.

Any changes made to the instrument communication settings will not be restored after a factory reset. See section

5.2 for more details.