5 electrical interface – Fluke Biomedical 956A-201-M2 User Manual

Page 21

Advertising
background image

2-3

4. The fourth mounting utilizes the Victoreen Model 948-10 Rack Chassis Adapter to mount a 956A

UDR into an existing Victoreen analog ratemeter rack chassis such as those used for the Victoreen
842 series analog ratemeter. Refer to drawing GEL-948-10 for mounting dimensions.


The UDR is simply inserted into the selected mounting enclosure and secured using the front panel pawl
fastener.

2.5 ELECTRICAL INTERFACE

Ensure all power is off prior to connecting the field wiring

Electrical interconnections at the detector are made via the mating "MS" style connector supplied with the
detector and loop drawing S174014A-102.

Electrical interconnections to the UDR are implemented with the mating connector kit, P/N 942A-100- 4
(supplied with the readout) by using the connector input/output assignments provided in Tables 2-3
through 2-10, and loop drawing S174014A-102, provided in Appendix B. Termination instructions for
coaxial connections can be found in Appendix D. A service loop of a minimum of 12 to 16 inches should
be provided to permit partial withdrawal of the UDR for setpoint adjustment and troubleshooting.

Connector pins should be soldered to the cable using 60/40-tin/lead resin core solder and a soldering iron
of 50 watts or less.

To protect the monitor from EMI/RFI transients, the mounting chassis must be earth grounded.

Table 2-3. Rear Panel Connections

Connector Function

P1

Relay/Output, Remote Alarm Acknowledge

P2 Detector

Connector

P3

AC Power Input

P4

Detector High Voltage (SHV)

P5

Detector Signal Input (BNC)

P6

Auxiliary outputs 0 – 10 Vdc & 4 – 20 mA DC

P7 RS232C,

Optional

WARNING

Advertising
This manual is related to the following products: