Fluke Biomedical 956A-201-M2 User Manual

Page 82

Advertising
background image

5-4

2.

If any voltage does not meet the tolerance after adjusting the +5.0 Vdc supply, the unit should be
returned to Cardinal Health, Radiation Management Services for replacement of the DC power
supply.

Discriminator Check, 956A-201 UDR

1.

Set the digital voltmeter to DC volts and measure the upper and lower discriminator voltages using
the values listed in Table 5-2.

Table 5-2. Discriminator Test Points

Test Point

Description

Expected Voltage

Potentiometer

GND Ground

N/A

N/A

LO Low

Discriminator

0.5 Vdc

± 0.01 V

VR11

HI High

Discriminator

7.000 Vdc

± 0.01 V

VR10

2.

If either voltage is out of tolerance, it may be brought into tolerance by adjusting the appropriate
potentiometer.

3.

If the proper adjustments cannot be achieved, return the unit to Cardinal Health, Radiation
Management Services for repair.

High Voltage Check, 956A-201 UDR

1. With the digital voltmeter set to read DC voltage, measure the detector high voltage as listed in Table

5-3.

Table 5-3. High Voltage Test Points

Test Point

Description

Expected Voltage

Potentiometer

GND Ground N/A

N/A

HV Test

HV, 1000:1

897A-210 0.575 V

± .01 Vdc

897A-220 0.575 V

± .01 Vdc

897A-230 0.550 V

± .01 Vdc

R5
R5
R5


2. If the 1000:1 voltage is out of tolerance, adjust R5 as required.

3. If the proper HV adjustment cannot be achieved, return the unit to Cardinal Health, Radiation

Management Services for repair.

Analog Output Check, 956A-201 UDR

1. To adjust the low scale output, enter 1.00E2 (a value greater than the background reading) into

setpoint 7, Analog Low Scale setpoint.

2. Adjust the appropriate zero adjust potentiometer as listed in Table 5-4 as required. The output signals

may be measured at the terminals as listed in Table 5-5.


Advertising
This manual is related to the following products: