3-32

Read-Write Cycles

A read cycle is performed by sequencing /RAM 0000 and /RAM READ, while /RAM WRITE is held high
(inactive). The address (A00 - A12) are latched by the failing edge of /RAM 0000. Data becomes valid
approximately 250 ns later.

A write cycle is performed by sequencing /RAM 0000 and /RAM WRITE, while holding /RAM READ high
(inactive). Identical to the read cycle, the address (A00 - A12) are latched by the falling edge of /RAM
0000. Data is stored into RAM on the failing edge of /RAM WRITE.

Write Register Decoding

Decoding for write registers within the main circuit board is performed by U2, which is a 1 of 8 decoder.
Control signals for U2 are /REGSELECT, /WRITE02, and /R/W (active high), as well as addresses A01,
A02, and A03. U2 decodes two addresses per output, starting at 4000, and ending with 400E. These
outputs are active low. Table 3-11 lists the write register assigned functions and addresses.

Table 3-11. Write Register Functions

ADDRESS STATE

A3 A2 A1

HEX

ADDRESS

FUNCTION

0 0 0 4000 Status

Indicators (/STATUS INDICATORS)

0 0 1 4002 E

2

PROM (/EEPROM)

0

1

0

4004

D/A Converter (/DAC)

0

1

1

4006

Display Control (/DISP. CONTROL)

1

0

0

4008

Display Data (/DISP. DATA)

1

0

1

400A

Counter Control (/COUNTER CONTROL)

1

1

0

400C

Relay Control (/400E)

1 1 1 400E

(Spare)

Counter Control (Write Only)

The counter control register (U43) is an 8-bit register with clear, and responds to address 400A. Upon
initial power-up (/RESET), the system reset signal sets all outputs low. The /COUNTER CONTROL
initiates data transfer to the register. Table 3-12 lists the counter control register write functions.

NOTE

Data written into U43 remains at the outputs until a reset occurs
or new data is written.

Table 3-12. Counter Control Register Write Functions

DATA BIT FUNCTION
D0 Counter

Clear

D1 Counter

Enable

D2 HV

Shutdown

D3 HV

Select

D4

Clear Switch Latch

D5 Not

Used

D6 Not

Used

D7 Not

Used

NOTE