Fluke Biomedical 956A-201-M2 User Manual

Page 60

Advertising
background image

3-34

The outputs of U5 drive the latch enable pins on U50-U59 in pairs as shown in Table 3-11. U50-U59 are
BCD-to-7 segment latch/decoder/drivers. Each drives five LEDs on the bargraph, either red or green, with
the exception of U54 and U59 that drive four LEDs each. BCD to bargraph decoding is accomplished by
latching data into the appropriate device to turn on the required bargraph segments. Segment decoding is
shown below:

Typical 7-Segment Digit


BCD Character

Segment(s)

1 c
7 c,

a

3 c,a,g
9

c, a, g, f

8

c, a, g, f, e

A-F None


Assume that all bargraph segments are off. Example: Write hex data 38 to bargraph address 4020. Data
38 is stored, low byte (8) in U55 and high byte (3) in U56. Data (8) in U55 causes the code for an 8 to be
output, which drives 5 output segments (c, a, g, f, e). The result is that the first five green LEDs turn on.
Data (3) in U56 causes the code for a 3 to drive segments c, a, and g connected to that device. The result
is that the first 3 of 5 segments associated with U56 will also turn on green.

Status Indicators (Write Only)

The status register (U60) is an 8-bit register with clear, and responds to address 4000. Data written into
U60 remains at the outputs until a reset occurs (/RESET), or new data is written (/STATUS INDICATORS).
Upon initial power-up, the system reset signal sets all outputs low. The status indicator write functions are
listed in Table 3-15.

Table 3-15. Status Indicator Write Functions

Data Bit

Function

Indication

D0

Check Source Indicator

D7

D1 Rate

D36

D2 Overrange

Indicator

D35

D3 Warn

Indicator

D33

D4

High Alarm Indicator

D32

D5 R/hr

Backlight

D6

D6 kR/h

Backlight

D4

D7 mR/h

Backlight

D5


U60 outputs, when high, control U61 inverter/driver to activate the appropriate front panel status
indicators. U60 outputs, when low, control U61 to deactivate the appropriate front panel status indicators.

LED 1 of 5 =

1

LED 2 of 5 =

7

LED 3 of 5 =

3

LED 4 of 5 =

9

LED 5 of 5 =

8

None =

A = F