Connector designations, Appendix a, A.1 connector designations – Fluke Biomedical 956A-201-M2 User Manual

Page 89

Advertising
background image

A-1

Appendix A

CONNECTOR

DESIGNATIONS

A.1 CONNECTOR DESIGNATIONS

Table A-1. Rear Panel Connections

Connector Function
P1

Relay/Output, Remote Alarm Acknowledge

P2 Detector

Connector

P3

AC Power Input

P4

Detector High Voltage (SHV)

P5

Detector Signal Input (BNC)

P6

Auxiliary Outputs; 0-10 VAC & 4-20 mA DC

P7 RS232C

Option


Table A-2. Connector P1 – Input/Output, 956A-201-M2

Pin Signal

Internal

Connection

1

120 Vac, Line

Relay Board K1-A/V2

2

120 Vac, Neutral

P2-13

3

120 Vac, Ground

P2-11

4 Spare

Relay

Board

K1-D

5 Spare

Relay

Board

K1-F

6 Spare

Relay

Board

K1-E

7

Fail Relay, common

Relay Board K3-A

8

Fail Relay, n.o.

Relay Board K3-B

9

Fail Relay, n.c.

Relay Board K3-C

10

Fail Relay, common

Relay Board K3-D

11

Fail Relay, n.o.

Relay Board K3-F

12

Fail Relay, n.c.

Relay Board K3-E

13

Warn Relay, common

Relay Board K4-A

14

Warn Relay, n.o.

Relay Board K4-B

15

Warn Relay, n.c.

Relay Board K4-C

16

Warn Relay, common

Relay Board K4-D

17

Warn Relay, n.o.

Relay Board K4-F

18

Warn Relay, n.c.

Relay Board K4-E

19

Alarm Relay, common

Relay Board K5-A

20

Alarm Relay, n.o.

Relay Board K5-B

21

Alarm Relay, n.c.

Relay Board K5-C

22 Spare

Not

Used

23 Spare

Not

Used

24 Spare

Not

Used

25

+ Remote Acknowledge

Main Circuit Board J9-1

26

- Remote Acknowledge

Main Circuit Board J9-2