5-10

High and Low Discriminator Adjustments

The low discriminator should be set to + 0.5 V +/- 0.01 Vdc, while the high discriminator should be set to
+7.0 +/- 0.01 Vdc unless otherwise specified in the detector calibration.

The following factory adjustments are included for general information and should not be performed in the
field.

Anti-jam Threshold Adjustment (Factory Adjustment Only)

This adjustment is dependent on the type of detector used. Potentiometer VR12 is used to adjust the DC
level measured at R94. Table 5-6 shows the DC adjustment values necessary for each detector type
unless otherwise specified.

Table 5-6. Anti- Jam Threshold Adjustment

Test Pulse Width

Detector Type

DC Value at R94 Equivalent mR/h

Jumper J7

10 microsecond
(50 kHz square wave)

897A Series
G-M Detector Family

0.740 V

897A-21X 3.16 x 10

3

897A-22X 2.14 x 10

4

897A-23X 2.50 x 10

5

2 to 3

Write Cycle Clock Adjustment (Factory Adjustment Only)

1. Connect the channel 1 oscilloscope (x10 probe) to U19-2 (clock).

2. Connect the channel 2 oscilloscope (x10 probe) to U19-12 (short 02).

3. Adjust VR13 to obtain: